Environmental Catalogue

BACK TO CATALOGUE

U-10 Multiparameter Water Quality Meter

 

U-10 MULTIPARAMETER WATER QUALITY METER
Meter for Monitoring pH, Conductivity, Dissolved Oxygen, Turbidity, Salinity, and Temperature
(NEW PRODUCT!)

Multiparameter water quality meter

FEATURES:

 

bullet

Rugged

bullet

Fully submersible sensor

bullet

Ideal for multi-parameter water quality measurements without having a permanent installation

multiparameter water quality meter and probe

On this page:

- Product Description
- Specifications
- Options & Prices

Additional Information:

- Product Support

Product Description

Global Water's U-10 Multiparameter water quality meter brings laboratory accuracy and the ease of push-button operation to water quality measurement in the field.  Six parameters can be measured : pH, temperature, dissolved oxygen, electrolytic conductivity, turbidity, and, salinity.  Featuring a powerful sensor and state-of-the-artfunctions, the Multiparameter water quality meter makes it easy to get highly reliable water quality data simply and quickly just by submersing the sensor in the water.  The Multiparameter water quality meter is flexible enough to use for cheeking the quality of a broad range of water samples, from factory effluent to urban drainage, river water, lake and marsh water, aquatic culture tanks, agricultural water supplies, and sea water.  The main unit is water resistant, so it can be used safely even in the rain.

Automatic Calibration

The U-10 Multiparameter water quality meter has an automatic calibration function that calibrates the instrument simply at the touch of a button for more highly reliable data.  What used to be a complicated calibration process is now easy.  Just fill the supplied beaker with the standard solution, immerse the sensor and press a button once for one-point calibration for each of the four parameters pH, conductivity, turbidity and DO.  Two-point calibration is also possible should more precise data be required.

Automatic Salinity Correction

The U-10 Multiparameter water quality meter has an automatic salinity function that makes it possible to measure DO in both fresh water and salt water.  It is now possible to measure DO in either fresh or salf water with the same instrument.  When measuring DO in salt water, it will automatically measure the salt concentration and make the salinity correction.  It is also possible to set a freely selectable salinity correction factor of between 0 and 4 percent.

Limited Data Logger

20 samples of data for all 6 parameters can be stored in memory to be recalled later for display and print out.  Store the results of measurement of all six parameters for up to 20 samples.  There is no need to record measured values on site.  Data stored in memory can be recalled and displayed at any time, and can even be simultaneously printed out on any standed printer by connecting it to the Centronics-specification printer output terminals.

The standard handheld multi-parameter water quality measurement system comes with:  

bullet

Carrying case with strap for handheld multi-parameter meter and water-quality probe

bullet

Handheld Multiparameter water quality meter with water quality probe

bullet

Instruction manual

bullet

pH 4 Buffer

bullet

Reference inner solution

bullet

Calibration beaker

bullet

DO installation tool

Specifications

 

PARAMETERS

pH

CONDUCTIVITY

TURBIDITY

DISSOLVED OXYGEN

TEMPERATURE

SALINITY

PRINCIPLE

Glass electrode

Alternating four-electrode

Scattering/transmitting light

Membrane/galvanic cell

Thermistor

(Alternating four-electrode)*

RANGE

0-14 pH

0-100 mS/cm

0-800 NTU

0-19.9 mg/l

0-50° C

0-4 %

RESOLUTION

STANDARD

0.1 pH

(mS/cm)  (mS/cm)
     0-1     :0.01
   1-10    :0.1
10-100  :1

10 NTU

0.1 mg/l

1° C

0.1%

EXPANDED

0.01 pH

(mS/cm)  (mS/cm)
     0-1    :0.001
   1-10    :0.01
10-100  :0.1

1 NTU

0.01 mg/l

0.1° C

0.01%

REPEATABILITY

+0.05 pH

+1% of full scale

+3% of full scale

+0.1 mg/l

+0.3° C

+0.1 %

TEMP. COMPENSATION

0-50° C

0-50° C

N/A

0-40° C

N/A

N/A

CALIBRATION

Auto one-point calibration, manual two-point calibration.

Auto one-point calibration, manual two-point calibration.

Auto one-point calibration, manual two-point calibration.

Auto one-point calibration (air), manual two-point calibration.

N/A

N/A

DISPLAY

LCD

PRINTER OUTPUT

Centronics

POWER

Battery 6F22(S-006P)x1pc.

SENSOR ASSEMBLY

Standard: sensor with 2 m cable. Optional: Sensor with 10 m cable.

WEIGHT

Main Unit: ~400g(0.9lbs), Sensor Assembly (2 m cable): ~800g(1.8lbs)

Note:
* Salinity is calculated from the conductivity data

Options & Prices

U-10 Multiparameter Probe
Probe includes 2 m cable.

$2,850

U-10 Multiparameter Probe (with 10 m cable)
Longer cable lengths available on request.  Specify cable length when ordering.

$3,200

U-10 Flow Cell $480
U-10 REPLACEMENT PARTS  
pH sensor $86
DO sensor $369
Reference sensor $80

RELATED PRODUCTS

U-22XD Multiparameter Probe
Use the multiparameter probe to monitor 10 different parameters, including total suspended solids,

water depth, and ORP.

$5320

W-21 SDI-12 Multiparameter Probe
Analyze eight water quality parameters with SDI-12 output.

$4070

Water quality sensors for permanent installation.

 

 
Up | Next
BACK TO SITE | BACK TO ENVIRONMENTAL CATALOGUE

 