Canada Lab Instruments Catalogue

Back To Catalogue

 

CO 2  INCUBATOR

bullet

AF 151UV (151 liters volume with UV disinfection)

bullet

AF 212UV (212 liters volume with UV disinfection)

bullet

HF 90(151 liters volume with 90 moist heat disinfection)

bullet HF 240(240 liters volume with 90 moist heat disinfection)

Precise CO2 Control

Based on the proven thermal-conductivity testing method, a reliable CO2 measuring sensor is installed to measure and control the concentration of CO2 gas in the chamber, which ensures long-term stability of PH-values, even in the lower CO2 range.

Our products are produced under total quality control and each unit is tested strictly before delivery. Therefore, our products are widely accepted by customers because of high performance, reliable quality and affordable price.

Easy and Reliable Operation

" Easy and reliable operation for standard applications "  is our customer's requirement for standard CO2 incubators. Shanghai Lishen is glad to offer four models of standard CO2 incubators as AF151UV, AF212UV, HF90 and HF240 to meet such kind of demand. From feedback we know, our incubators will bring the customers full satisfactions.                 

Microprocessor Control

All control functions and data are controlled and monitored by a microprocessor. If a failure happens, the microprocessor will give out warning messages and "Error Code" messages.

 

 

 

 

                   AF151UV

Heating System

Unique direct heating and air jacket system maintain the inner chamber at a constant and stable temperature. This allows our incubators to operate under a perfect temperature / humidity environment and to maintain appropriate osmotic pressure in cells. The inner chamber is surrounded by a layer of insulation material, which prevents the incubator's temperature from dropping too fast during a certain case of power failure. The out door of the unit is also heated and the heating can be switched off separately if necessary. Also, it eliminates condensation on the interior glass door.

 

Reliable Temperature Control

High grade temperature sensor made of platinum (Pt1000), together with the micro-processor controller ensures accurate and reliable temperature measurement and control. Thus, the incubators can offer a reliable temperature control within the control range from RT+5 to 50. Even more, an independent temperature sensor for over-temperature protection is installed for double controlling. It will take over the temperature controlling function if the main temperature controller fails.

 

High Humidity

The incubators operate at a high relative humidity. This protects samples from drying out and maintains the appropriate osmotic pressure in cells, which is ideal for long-term cultures.

"Auto-start" Function

Through "auto-start" function, the incubators will set up an accurate culture environment when the incubators are turned on for operation.

UV Disinfection (for AF151UV&AF212UV only)

A long life UV light is installed in the chamber to optimize contamination elimination.

Interior Design

Smooth and corrosion-resistant stainless steel surface with special designed round inner corners makes cleaning work easier.

 

Rapid Recovery Time

It ensures that the culture conditions in the chamber recover quickly, which is very important for short-term cultures, or when the door is frequently opened and closed.

 

Special Advantages of AF151UV&AF212UV

bullet

"Auto-start" function

bullet

Microprocessor controlling system

bullet

Precise CO2 control

bullet

Independent over-temperature protection with

A second sensor

bullet

UV light for chamber disinfection

bullet

Easy operation with soft-touch control panel

bullet

Heating of out door can be switched on/off

HF 90 and HF240

Our new generation of CO2 Incubators

                          HF90

90 Moist Heat Disinfection

HF90 and HF240 are equipped with a new function called 90moist heat disinfection. By pressing the button of90 on the control panel, the incubator will carry on a complete circle of disinfection procedure. The incubator will increase the chamber temperature up to 90 and maintain it for 9 hours. The humidity inside the chamber will be above 80%. Then it will recover to the working parameters in about 14 hours (depending on the environment temperature). All the accessories inside the chamber like the CO2 sensor, fan and shelves are heat-resistant. The customer just need to take out the culture flasks or dishes, and the disinfection procedure will be done automatically.

 

 

                                HF240

The results of 90 Moist Heat Disinfection procedure are tested strictly according to international standards. The germ, Bacillus subtilis varietas niger spore (ATCC 9372), which is recommended for official sterilization by Ph. Eur. 97, 5.1.2 and USP 23. (3), is employed to test the result of 90_ Moist Heat Disinfection function. In average, 5.5 x 105 cfu/sample is distributed on all the surfaces of the chamber. After the disinfection procedure, all the sample areas are tested as negative and the elimination rate is 100%. The result is approved by Chinese Government Laboratory (see Form 1 as follows).

During daily applications, one of the most headache problems is called microorganism contamination inside the incubator chamber. To eliminate the contamination, normally people use manual cleaning, UV light disinfection (as AF151UV and AF 212UV), or HEPA filter filtration. However, none of them can offer a complete solution to eliminate contamination. Now, HF90 and HF240 CO2 incubators can be a new solution.

Form 1. Elimination Rate by Disinfection Function of HF 90

Test germ: Bacillus subtilis varietas niger spore

(ATCC 9372).

3 - pieces Inner Glass Door (only for HF90)

 

            HF90 (3-pieces inner glass door)

 

HF 90 CO2 incubator is equipped with a set of 3-pieces inner glass door as standard offer in each unit. It means that besides the one piece of interior glass door, there is still a set of 3-pieces inner glass door, just opposite to 3 shelves inside the chamber.

It can greatly help the customer to reduce influence to the culture environment inside the chamber when the door is opened frequently. At least, two thirds of the door area will not be opened if the customer is just moving flasks or dishes on one shelf. Then, less temperature and gas will be lost, After closing the door; the culture environment will recover back to setting points very fast.

HF90 and HF240 have other features as AF151UV and AF212UV:

bullet

Microprocessor control

bullet

Precise CO2 control

bullet

Direct heating and air jacket system

bullet

Reliable temperature control

bullet

High humidity

bullet

"Auto-start" function

bullet

Independent over - temperature protection

With a second sensor

bullet

Easy operation with soft-touch control panel

bulletHeating of out door can be switched on/off

 

                      Control Panel for HF90

 

 

 

Technical Specifications

Items

AF151UV

AF212UV

HF90

HF240

Dimensions

 

 

 

 

Outer casing (wxhxd mm)

615x865x680

910x795x763

637x909x762

780x944x820

Working chamber(wxhxd mm)

470x607x530

600x600x588

470x607x530

607x670x583

Inner volume (liter)

151

212

151

240

Shelves (wxd mm)

423x445

550x510

423x445

554x503

No. of shelves(standard pcs)

3

3

3

3

No. of shelves(maximum pcs)

10

10

10

12

Temperature

 

 

 

 

Heating system (℃)

Direct heating

Direct heating

Direct heating

Direct heating

Control range (℃)

RT+5~50

RT+5~50

RT+5~50

RT+5~50

Deviation at 37℃ (Temporal* ℃)

±0.1

±0.1

±0.1

±0.1

               (Spatial* ℃)

±0.4

±0.4

±0.4

±0.4

CO2 system

 

 

 

 

Measurement sensor

TCD

TCD

TCD

TCD

Control range (%)

0~20

0~20

0~20

0~20

Deviation at 37℃ (%)

±0.1

±0.1

±0.1

±0.1

Humidity

 

 

 

 

Relative humidity

≥95%

≥95%

≥95%

≥95%

Distilled water volume (Max. liter)

4

4

3

3

Disinfection

 

 

 

 

Disinfection method

UV light

UV light

90℃ moist heat

90℃ moist heat

Disinfection duration

30 minutes

30 minutes

25 hours

25 hours

Electric data

 

 

 

 

Rated voltage (V/Hz)

220/50

220/50

220/50

220/50

Connected load (KW)

0.6

0.77

0.6

0.735

Weight (kg)

75

95

80

98

* According. to DIN12880, part 2                           

 

 

 

BACK TO SITE | BACK TO CATALOGUE